Nuicon
[email protected]
Facebook
Twitter
Youtube
Linkedin
Home
Who We Are
About us
Ethics and Values
QHSE
Products
Sensors
Instrumentation
Automation
Our Services
Calibration and Maintenance Services
Contact
Home
Who We Are
About us
Ethics and Values
QHSE
Products
Sensors
Instrumentation
Automation
Our Services
Calibration and Maintenance Services
Contact
Request a Quote
High-precision pressure sensor with flush diaphragm and touch screen
CombiPress PFMN
Fully welded process connection
Pressure measuring ranges from -1 ... 0 bar to 0 ... 400 bar
Precision class ≤ 0.1% FS
Absolute pressure, relative pressure and vacuum measurement
CIP-capable
Optional switching output
Technical Data
Software
Accessories
Technical Specification
Measuring range
-1 … 400 bar
Min. measuring span
0.05 bar
Max. measuring span
401 bar
Pressure type
Absolute
Relative (gauged)
Standard error of measurement (BFSL)
0.04 % FSR , up to 2:1 turndown ratio
0.1 % FSR , up to 4:1 turndown ratio
Including non-linearity, hysteresis and non-repeatability according BFSL
For turndown, multiply this value by the applied turndown ratio
Max. measuring error
± 0.1 % FSR , up to 2:1 turndown ratio
± 0.25 % FSR , up to 4:1 turndown ratio
Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2)
For turndown, multiply this value by the applied turndown ratio
Temperature coefficient
≤ 0.05 % FSR/10 K , measuring span
≤ 0.05 % FSR/10 K , zero point
Compensated temperature range
-40 … 85 °C
Long term stability
≤ 0.1 % FSR/a , IEC 770 6.3.2
Max. turndown ratio
10 : 1
Rise time (10 ... 90 %)
≤ 0.3 s
Sample time
≤ 0.3 s
Power-up time
< 10 s